Chapter 13 Specifications
258
(1) Resistance range switching function
(2) 100 M range high-precision function
(3) Number of measurement digits selection function
(4) Low-Power Resistance Measurement function (LP)
(5) Measurement Current Switching
Functions
Mode AUTO/ MANUAL
(Automatically set to manual if the comparator or BIN function is turned on.)
Measurement Ranges LP OFF:
10 m/ 100 m/ 1000 m/ 10 / 100 / 1000 / 10 k/ 100 k/ 1000 k/
10 M/ 100 M/ 1000 M
LP ON: 1000 m/ 10 / 100 / 1000
(With the 100 M range high-precision setting on, the 1,000 M range cannot
be used. When using the MUX measurement terminal setting with the 2-wire
measurement method, the 10 and lower ranges cannot be used.)
Default setting Mode: AUTO, Measurement Range: 1000 M
Setting ON/ OFF
Default setting OFF
Number of measure-
ment digits selection
7digits/ 6digits/ 5digits (If the number of f.s. digits is less than the setting, the
number of f.s. digits will be used.)
Default setting 7digits
Operation Low-power measurement is performed by limiting the measurement current
and open voltage. (1000 m to 1000 range)
Setting ON/ OFF
(With OVC on when LP is on and the contact improvement function fixed to off)
Default setting OFF
Operation The measurement current is limited during measurement.
(100 m to 100 range)
Measurement current High/ Low
Default setting High
Range
Measurement current
High Low
100 m 1 A 100 mA
1000 m 100 mA 10 mA
10 10 mA 1 mA
100 10 mA 1 mA