10.4 External I/O Settings
209
10
ON edge and OFF edge operation
ON edge
OFF edge
Measurement
TRIG
EOM
ON OFF
OFF
Measurement
TRIG
OFF
OFF
Measurement
processing
Measurement
processing
EOM
Measurement times will be 0.2 ms longer when using the OFF edge than when using
the ON edge.
ON
ON
ON