4.3 Zero Adjustment
70
2
Verify that the measured value is within ±1%f.s. If the measured value is
50%f.s. or less in each range, zero-adjustment can be performed, but a warn-
ing will be issued when it is greater than 1%f.s.
If no measured value is displayed, verify whether the measurement leads
have been wired properly.
3
Perform zero-adjustment.
4
A confirmation message will be displayed. Confirm and return to the Mea-
surement screen.
Proper wiring Improper wiring
[0ADJ]
Perform zero-adjustment.
Switch the function menu
to P.2/3.
1
2
Perform zero-adjustment
and return to the Measure-
ment screen.
Cancel the operation and
return to the previous
screen.