Trident HD Specimen Radiography System User Guide
Appendix A: System Specifications
Page 74 MAN-05478-002 Revision 003
A.2 Operation and Storage Environment
A.2.1 General Conditions for Operation
This equipment is designed for safe and effective operation under the following
conditions for Indoor Use:
Altitude
To 3,048 m (10,000 feet)
Degree of Pollution
Pollution Degree 2
Temperature Range
15 °C (59 °F) to 30 °C (86 °F)
Relative Humidity Range
20% to 80% without condensing moisture
A.2.2 General Conditions for Transport and Storage
Console
Temperature Range
10 °C (14 °F) to 40 °C (104 °F) after being stored for 48
hours
10 °C (50 °F) to 35 °C (95 °F) for up to 12 hours
Maximum rate of temperature change
Less than 10 °C (50 °F) per hour
Relative Humidity Range
20% to 90% without condensing moisture
X-ray Detector
Temperature Range
10 °C (50 °F) to 30 °C (86°F) after being stored for 48 hours
10 °C (50 °F) to 35°C (95 °F) for up to 12 hours
Maximum rate of temperature change
Less than 10 °C (50 °F) per hour
Relative Humidity Range
20% to 90% without condensing moisture
(Put in a package for storage in a building.)