Chapter 2: General Information
Page 8 MAN-02098 Revision 008
Caution:
The weight and mobility of this system require the
following precautions:
Be careful when you move the system on uneven or
sloped surfaces.
Make sure the wheels are locked when the system is
stationary.
Before you move the cart, make sure the cables are
stored in a safe position on the cart.
Caution:
The system is a laboratory device and not a normal
computer. Do not make changes to the hardware or
software that are not authorized. Install this device behind
a firewall for network security. The computer virus
protection or network security for this laboratory device is
not provided (for example, a computer firewall). The
network security and anti-virus provisions are the
responsibility of the user.
2.4 Interlocks
The x-ray generation stops automatically and an error message appears when there is an
interruption of an x-ray exposure such as when the mAs timer or maximum exposure
timer times out, the cabinet door is opened, or the key switch is turned to the locked
position. To make another exposure:
1. Select the OK button in the error message dialog box.
2. Select the status icon (shaped like the Trident unit) in the taskbar at the bottom of the
screen.
3. Complete any action (such as "Close the door") that shows in the system message area
of the screen.
4. Select the Clear All Faults option.
5. When READY shows in the system message area, press the X-RAY button when you
are ready to take another exposure.