Appendix A: System Specifications
Page 54 MAN-02098 Revision 008
A.3 Workstation Technical Information
A.3.1 General Information
Operating System
Windows 7
Hard Disk Drive Capacity
500 GB minimum
Storage Device Disks
DVD+/-RW
Display
1920 x 1200 pixel resolution
A.3.2 Network Environment
Network Interface
Integrated Gigabit Ethernet controller
Cable Connection
RJ45
Cable Category
5e (minimum)
Cable Length
3 meters (maximum)
A.4 Electrical Input
The system is designed for temporary overvoltages on the Mains supply.
Line Voltage Range
100-240 VAC ±10%
Mains supply voltage fluctuations
To 10% of the nominal voltage
Transient overvoltages
To the levels of Overvoltage Category II
Operating Frequency
50/60 Hz ±2%
Peak Line Current
4 amps
Power Consumption
450 watts maximum
Line Connection
Standard line cord and plug - medical grade
UPS Operation
The (optional) UPS provides full power to the system for 30 minutes
from a full charge (limited to four exposures within the 30 minutes)
Duty Cycle
10 seconds on, 20 seconds off
A.5 X-ray Tube Technical Information
Source-to-Image Distance (SID)
49 cm nominal
Focal Spot
50 micron nominal
Tube Voltage
20 kV to 35 kV
Anode Material
Tungsten
X-Ray Window
Beryllium 125 µm
A.5.1 X-Ray Generator
kV Range
20 kV to 35 kV
mA
1 mA
mAs
10 mAs maximum