Trident User Guide
Appendix A: System Specifications
MAN-02098 Revision 008 Page 55
A.6 Imaging System Technical Information
A.6.1 Image Receptor
Fluid Ingress
The Image Receptor is protected from accidental spillage on the Cabinet
floor.
Active Imaging Area
12 cm by 14 cm nominal
Dynamic Range and Linearity
Detector Subsystem response is linear with linearity of 0.999 over a
dynamic range of 400:1 in x-ray exposure.
Uniformity
Detector Subsystem can correct pixel-to-pixel gain variations.
Image Pixel Size 70 µm
Digital Image Receptor MTF
> 40% at Nyquist frequency