HOUSING INTERIOR WIT
H MAJOR INTERNAL COM
PONENTS
EnCal 3000
Hardware
Manual
73022344 o
38
Detector
Solenoid V1
Solenoid V2
Solenoid V3
Solenoid V4
Solenoid V5
inlet
Solenoid V6
outlet
sample
loop
inject valve
Manifold block
Column vent
Reference vent
Sample vent
EGC/Ref vent
Sample
Injector
Analytical
column
Reference
column
Pump Manifold
Pressure
sensor
Pre
column
Pump
sample valve
Pump valve
Backflush
valve
BF vent
Detector
Solenoid V1
Solenoid V2
Solenoid V3
Solenoid V4
Solenoid V5
inlet
Solenoid V6
outlet
sample
loop
inject valve
sample valve
Manifold block
Column vent
Reference vent
Sample vent
EGC/Ref vent
Sample
Injector
Analytical
column
Reference
column
Pump Manifold
Pressure
sensor
Pre
column
Backflush
valve
Pump
Pump valve
BF vent
Figure 5.4: Internal Gas Flow Circuit for each Analytical Channel with backflush of the
EnCal 3000, before activating backflush valve (top) and after activating backflush
valve (below)