95 ST 800 Pressure Transmitter User’s Manual Revision 15
Figure 27 Meter Body Reassembly
16. Use a torque wrench to gradually tighten nuts to torque rating in sequence shown in Figure .
Tighten head bolts in stages of 1/3 full torque, 2/3 full torque, and then full torque.
Figure 28 Head Bolt Tightening Sequence
17. Feed the ribbon cable on the new meter body through the neck of the housing.
CAUTION: To prevent damage to the ribbon cable, use care when assembling the meter
body to the electronics housing.
18. Screw the new meter body into the housing until the bottom of the meter body adapter is flush
with the neck of the electronics housing.
19. Tighten the outside set screw to be sure it is fully seated in the slot in the header.
20. Loosen the set screw ½- turn.
21. Rotate the housing to the desired position (Max. 180
o
in either direction) or tighten the set screw.