59 ST 800 Pressure Transmitter User’s Manual Revision 15
2ВвдLRV
[ 2ENTLRV]
3ВвдURV
[ 3ENTURV]
The limits are:
2X the Lower Range Limit
(LRL) of the Meter body and 2X
the Upper Range Limit (URL) of
the Meter body.
The LRV/URV value will be
available in Pressure Units
(Simple Display pressure unit)
Press to
enter menu
selection
Menu to
select
number.
to enter and
shift to the
next digit to
the right
perform an input calibration
correction by applying process
input at zero.
Press to
enter menu
selection
Press to
initiate action
5НолЦАП
[5ZEROTR]
Note: Loop must be
removed from
Automatic Control
This selection allows the loop
zero output 4mA value to be
trimmed.
Note: You must connect a
current meter to the
transmitter to monitor the loop
output.
Press to
enter menu
selection
Menu to
select
number.
to enter and
shift to the
next digit to
the right
6ШклЦАП
[6SPANTR]
Note: Loop must be
removed from
Automatic Control
This selection allows the loop
span output 20mA value to be
trimmed.
Note: You must connect a
current meter to the
transmitter to monitor the loop
output.
7ТокТст
[7 LPTEST]
Note: Loop must be
removed from
Automatic Control
This selection allows the user
to force the DAC output to any
value between 3.8 and 20.8
mA.
Note: This selection will put
the DAC into Fixed Output
Mode, as indicated by the
flashing output value.
Navigation away from this
menu item will return the loop
to Normal (Automatic) Mode.