4 Maintaining the MSD
To replace the diffusion pump fluid
117
5 Wipe clean the diffusion pump flange on the vacuum manifold.
Follow the instructions on the bottle for pre-heating the diffusion pump fluid.
6 Pour new diffusion pump fluid into diffusion pump until the fluid level is
within the
FULL COLD range.
It will require approximately 1.5 to 2 containers (18.5 ml each) of diffusion pump
fluid. Watch the sight glass while pouring. Do not overfill.
7 Reinstall the diffusion pump (page 118).