60
2 Operating the MSD
To connect the optional gauge controller
To connect the optional gauge controller
Materials needed:
Gauge controller (HP 59864B)
Power cord
Triode gauge cable (8120-6573)
1 Connect the triode gauge cable to the triode gauge tube.
2 Connect the other end of the triode gauge cable to the gauge controller.
3 Connect the power cord to the gauge controller.
4 Connect the other end of the power cord to an appropriate electrical
outlet.
If you wish to share one controller among MSDs, obtain one cable for each instru-
ment. Leave a cable connected to the triode gauge tube on each MSD. This will
avoid having to vent the MSD before connecting the controller.
: $ 5 1 , 1 * 1HYHUFRQQ HFW RUGLVFRQQ HFWWKH FDEOH IURPWKH W ULRGHJDXJH WXEHZKLOHWK H0 6'LVXQGHU
YDFXXP5LVNRILPSORVLRQDQGLQMXU\GXH WREURNHQJODVVH[LVWV
&$87,21
Do QRW use an HP 59864A (older model) triode gauge controller during data acquisition. This
model can be used for diagnostic purposes
RQO\
.
&$87,21
Be sure to orient the cable and the gauge tube as illustrated. Excessive force on the pins
can break the tube.