HP 5973 - Turbomolecular Pump; Foreline Pressure Has Exceeded 300 Mtorr; Internal MS Communication Fault; Lens Supply Fault

HP 5973
297 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
86
3 Troubleshooting the MSD
Error messages
Turbomolecular pump
Large vacuum leak is preventing the turbo pump from reaching 50% of full speed
Foreline pump is not working correctly
Turbo pump is not working correctly
Turbo pump controller is not working correctly
MSD electronics are not working correctly
Foreline pressure has exceeded 300 mTorr
Only diffusion pump-equipped MSDs have a foreline pressure gauge.
Excessive carrier gas flow (typically > 5 ml/min)
Excessive solvent volume injected
Large vacuum leak
Collapsed or kinked foreline hose
Foreline pump is not working correctly
Foreline gauge is not working correctly
Internal MS communication fault
MSD electronics are not working correctly
Lens supply fault
Electrical short in the analyzer
MSD electronics are not working correctly
Log amplifier ADC error
MSD electronics are not working correctly
No peaks found
Emission current was set to 0
Electron multiplier voltage is too low
Amu gain or offset is too high
Calibration valve is empty