76
3 Troubleshooting the MSD
Mass spectral symptoms
Mass assignments are incorrect
Small shape changes at the top of the mass peaks can cause 0.1 amu shifts in mass
assignments. Shifts greater than 0.2 amu indicate a possible malfunction.
MSD has not had enough time to reach thermal equilibrium
Large variations in the temperature of the laboratory
MSD has not been tuned recently, or at the temperature at which it is operating
Incorrect tune file (inappropriate parameters)
Peaks have precursors
The tune report lists the size of the precursors for the tune masses. Small precur-
sors are not unusual. There are no specifications for acceptable or unacceptable
precursor size. If the precursors are unacceptably large for your application, one
of the following may be responsible:
Repeller voltage is too high
Peaks are too wide
Incorrect dc polarity on the quadrupole mass filter
Dirty quadrupole mass filter
Peak widths are inconsistent
MSD has not had enough time to reach thermal equilibrium
Large variations in the temperature of the laboratory
Incorrect tuning
Calibration vial is empty or almost empty
Calibration valve is not working correctly
Dirty ion source
Electron multiplier is nearing the end of it useful lifetime
Ground loop