HP 5973 - To Lubricate the Vent Valve O-Ring

HP 5973
297 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
142
4 Maintaining the MSD
To lubricate the vent valve O-ring
To lubricate the vent valve O-ring
Materials needed:
Cloths, clean (05980-60051)
Gloves, clean, lint-free
large (8650-0030)
small (8650-0029)
Grease, Apiezon L, high vacuum (6040-0289)
O-ring, vent valve (0905-1217) - replace if the old O-ring is worn or damaged
The vent valve O-ring needs a thin coat of lubrication to ensure a good vacuum
seal and smooth operation. If the vent valve O-ring does not turn smoothly, or
does not seal correctly, lubricate it using this procedure.
1 Vent the MSD (page 52).
2 Completely remove the vent valve knob.
3 Inspect the O-ring.
If the O-ring appears damaged, replace it.
4 Use a clean, lint-free cloth or glove to spread a thin coat of high vacuum
grease on the exposed surface of the O-ring.
5 Use a clean, lint-free cloth or glove to wipe away excess grease.
If the O-ring looks shiny, there is too much grease on it.
&$87,21
Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be
lubricated. Lubricating other seals can interfere with their correct function.
&$87,21
Excess grease can trap air and dirt. Grease on surfaces of the O-ring other than the exposed
surface can trap air, resulting in air spikes during operation.