3 Troubleshooting the MSD
Mass spectral symptoms
75
Mass spectral symptoms
This section describes symptoms you might observe in mass spectra. Some of
these symptoms will appear in the mass spectra of samples. Others you will
observe only in a tune report. Some of these symptoms have causes that can be
corrected by the operator. Others, however, require service by a Hewlett-Packard
service representative.
Two symptoms listed under Chromatographic symptoms: If sensitivity is poor
and If repeatability is poor, also apply to mass spectra.
Isotopes are missing or isotope ratios are incorrect
Peaks are too wide or too narrow
Scan speed is too high (scan mode)
Dwell time is too short (SIM mode)
Electron multiplier voltage is too high
Repeller voltage is too high
High background
Dirty ion source
High background
Pressure in the vacuum manifold is too high
Air leak
•Contamination
High abundances at m/z 18, 28, 32, and 44 or at m/z 14 and 16
System was recently vented (residual air and water)
Air leak
Large peaks at m/z 14 and 16 are symptomatic of especially large leaks.