3 Troubleshooting the MSD
Error messages
The system is in vent state
Although the message says the system is in vent state, if the fault has just occurred
the MSD may actually still be under vacuum and the diffusion pump may still be
hot or the turbo pump may still be at high speed. Wait at least 30 minutes after
seeing this message before you actually vent the MSD.
System was vented on purpose (no problem)
Fan fault has turned off the high vacuum pump (power cycle the MSD to clear
the fault)
Fuse for the high vacuum pump has failed
Diffusion pump heater cartridge has failed
MSD electronics are not working correctly
There is no emission current
Filament is not connected properly
Filament has failed
MSD electronics are not working correctly
There is not enough signal to begin tune
Poor mass axis calibration
Amu gain or offset is too high
Calibration vial is empty or almost empty
Excessive pressure in the vacuum manifold
Air leak
Electron multiplier voltage is too low
Signal cable is not connected
Electrical leads to the detector are not connected correctly
Electrical leads to the ion source are not connected correctly
Filament shorted to the source body
&$87,21
Venting the MSD too soon after this message appears can result in the backstreaming of
a diffusion pump or damage to a turbomolecular pump.