4 Maintaining the MSD
To remove the calibration vial
To remove the calibration vial
Materials needed:
None
1 Stop any tuning or data acquisition.
2 Turn off the analyzer.
There are several ways to turn off the analyzer:
In the Diagnostics/Vacuum Control view, select
MS OFF from the Diagnostics
menu.
In the Instrument Control view in the Edit Parameters dialog box, select
MS OFF
from the Execute menu.
In the Manual Tune view, select
MS OFF from the Execute menu.
Type
MSOFF on the command line and press Enter.
Verify that only the power LED is lit on the front of the MSD.
3 If your MSD is equipped with a gauge controller, switch off the triode
gauge and the gauge controller.
4 Remove the upper MSD cover (page 50).
5 Loosen the calibration vial collar by turning it counterclockwise.
Counterclockwise as viewed from the bottom (vial side) of the collar. Do not
remove the collar.
6 Pull the calibration vial out.
You may feel some resistance due to residual vacuum.