3 Troubleshooting the MSD
Chromatographic symptoms
71
Does not pass tune
• Calibration vial is empty
• Excessive foreline or vacuum manifold pressure
• Very dirty ion source
• Calibration valve is not working correctly
• Bad signal cable connection
• Filament has failed or is not connected correctly
• Bad ion source wiring connection
• Bad detector wiring connection
• Failed electron multiplier horn
Peaks are tailing
• Active sites in the sample path
• Injection is too large
• Incorrect injection port temperature
• Insufficient column flow
• GC/MSD interface temperature is too low
• Ion source temperature is too low
Peaks are fronting
• Column film thickness mismatched with analyte concentration
(column overload)
• Initial oven temperature is too low
• Active sites in the sample path
• Injection is too large
• Injection port pressure too high
• Insufficient column flow