4 Maintaining the MSD
To remove the triode gauge tube
To remove the triode gauge tube
Materials needed:
Gloves, clean, lint-free
large (8650-0030)
small (8650-0029)
1 Vent the MSD (page 52).
2 Disconnect the cable from the triode gauge tube.
3 Loosen the triode gauge collar by turning it counterclockwise.
Do not remove the collar.
4 Pull the triode gauge tube out of the collar.
5 Remove the baffle from the open end of the triode gauge tube.
Wear clean gloves when handling the baffle. If you set the baffle down, make sure
it is on a clean surface.
: $ 5 1 , 1 * 1HYHUFRQQ HFW RUGLVFRQQ HFWWKH FDEOH IURPWKH W ULRGHJDXJH WXEHZKLOHWK H0 6'LVXQGHU
YDFXXP7KHVWUHVVFRXOGFDXVHWKHWX EHW RLPSORGH