4 Maintaining the MSD
To clean the ion source
153
8 Dry the cleaned parts in an oven at 100 ° – 150 °C for 30 minutes.
Take care to avoid recontaminating cleaned and dried parts. Put on new, clean
gloves before handling the parts. Do not set the cleaned parts on a dirty surface.
Set them only on clean, lint-free cloths.