4
The HP 5973 Mass Selective Detector
The HP 5973 MSD is a stand-alone capillary GC detector
The HP 5973 Mass Selective Detector (MSD) is designed for use with the
HP 6890 Series Gas Chromatograph. The MSD features:
Either a 90 L/sec vapor-diffusion or a 250 L/sec turbomolecular
high vacuum pump
Rotary vane foreline pump
Independently heated electron-ionization ion source
Independently heated hyperbolic quadrupole mass filter
High-energy dynode (HED) electron multiplier detector
Independently heated GC/MSD interface
Power supplies and instrument control electronics
Operator control of the MSD is through the data system
The data system features Hewlett-Packard MSD ChemStation software.
It includes programs to calibrate (tune) the MSD, acquire data, and analyze
data.
Tuning programs adjust voltages in the ion source, calibrate mass assign-
ments, and control the scanning of the analyzer.
Data acquisition programs allow you to monitor the total ion chromato-
gram (TIC), automatically storing the mass spectra of GC peaks as they
elute. This is the scan mode. Alternatively, you can monitor the concentra-
tions of particular ions. This is selected ion monitoring mode (SIM).
Data analysis programs allow you to integrate chromatograms (TIC or
SIM), view and manipulate mass spectral data, compare spectra to databas-
es of reference spectra in a library search, quantitate data, and generate
reports.