140
Chapter 6, Call Processing Subsystem
CALL DATA Screen
The messages which can be viewed are:
• Reverse Control Channel Messages for Paging, Origination, Orders, and Order Confir-
mation.
RECCW A - Word A - Abbreviated Address Word
RECCW B - Word B - Extended Address Word
RECCW C - Word C - Serial Number Word
RECCW D - Word D - First Word of the Called-Address
RECCW E - Word E - Second Word of the Called-Address
• Reverse Voice Channel Messages for Order Confirmation.
RVCOrdCon - Reverse Voice Channel Order Confirmation Message
The field descriptions for each of the decoded messages are given in the "CALL
DATA Screen Message Field Descriptions" on page 144
.
When the CALL DATA screen is displayed and the Call Processing Subsystem is
in the connected state, the Test Set constantly monitors the mobile station’s
transmitted carrier power. If the power falls below 0.0005Watts, the error
message
RF Power Loss indicates loss of Voice Channel will be
displayed and the Test Set will terminate the call and return to the active state.
NOTE: In order to ensure that the Test Set makes the correct decisions regarding the presence of the
mobile stations’s RF carrier, the Test Set’s RF power meter should be zeroed before using the
Call Processing Subsystem. Failure to zero the power meter can result in erroneous RF power
measurements. See "Conditioning the Test Set for Call Processing" on page 130 for
information on zeroing the RF Power meter manually.