EasyManuals Logo

Inficon XTC/3 User Manual

Inficon XTC/3
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
5 - 9
XTC/3 Operating Manual
Response Error Code = 1 byte ASCII
A= Illegal Command
B = Illegal parameter value
C = Illegal ID
D = Illegal format
E = Data not available
F = Cannot do now (some commands require XTC/3 to be
in Ready/Stop).
L = Length Error, must be greater than 0 and equal to or
less than 57,800.
O = Not enough room for response size.
P = Prior command failed (if one command of a multiple command
packet fails, none of the following commands will be done, and this
error code will be returned).
Packet Error Code = 1 byte ASCII.
C – invalid checksum
F – illegal format (too many bytes for the commands requested or
missing value in command)
I – invalid message
L – Length Error, must be greater than 0 and equal to or less
than 57,800.
M – too many commands (Only 100 allowed)
Checksum . . . . . . . . . . . . . . . . . . . . 1 byte, sum, modulo 256, of all bytes,
including CCB, Timer, and Response
Message but not including length.
Timeouts. . . . . . . . . . . . . . . . . . . . . . If more than 3 seconds pass between
characters of a command packet, XTC/3
will time out. No response packet is sent,
XTC/3 will clear its buffers and assume any
future characters are the beginning of a
new packet.

Table of Contents