4 - 5
XTC/3 Operating Manual
4.2.3 Film Definition Parameters
The Film parameters are entered in the Pre/Post, Deposit, Sensor, Source and
Option screens.
4.2.3.1 Pre/Post Deposition Parameters
RISE TIME 1 . . . . . . . . . . . . . . . . . . 00:00 to 99:59 min:s
This parameter provides the time period over which the source power
is ramped from 0 to SOAK POWER 1. The default value is 00:00.
SOAK POWER 1 . . . . . . . . . . . . . . . 0.0 to 100%
This parameter is usually set to the power level at which the source material
just begins to melt. XTC/3 ramps the power level from zero to
SOAK POWER 1 linearly over the time period RISE TIME 1.
The default value is 0%.
SOAK TIME 1 . . . . . . . . . . . . . . . . . 00:00 to 99:59 min:s
This parameter provides the time period for which XTC/3 holds at
SOAK POWER 1. The default value is 00:00.
RISE TIME 2 . . . . . . . . . . . . . . . . . . 00:00 to 99:59 min:s
This parameter provides the time period over which the source power is
ramped to SOAK POWER 2. The default value is 00:00.
SOAK POWER 2 . . . . . . . . . . . . . . . 0.0 to 100%
This parameter is usually set to the power level at which the source material
evaporates material at about the target rate. XTC/3 ramps the power level from
SOAK POWER 1 to SOAK POWER 2 linearly over the time period
RISE TIME 2. The default value is 0%.
SOAK TIME 2 . . . . . . . . . . . . . . . . . 00:00 to 99:59 min:s
This parameter provides the time period for which XTC/3 holds at
SOAK POWER 2. The default value is 00:00.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon XTC/3 and is the answer not in the manual?

Inficon XTC/3 Specifications

General IconGeneral
ModelXTC/3
Input Voltage100-240 VAC, 50/60 Hz
Operating Temperature0 to 50 °C
Rate Resolution0.01 Å/s
Measurement PrincipleQuartz Crystal Microbalance (QCM)
Power SupplyInternal
Communication InterfaceRS-232, Ethernet
DimensionsStandard 19" rack mount
WeightApprox. 5 kg
Crystal Sensor6 MHz Quartz Crystal
OutputsAnalog, Digital

Related product manuals