5.2.2.4.3 Standard Update Film Parameters
Command:
UF <Command ID> <Film Number> <Parameter Value>
Command ID = <Byte>. See “UF Command ID” column of Table 5-2.
Film Number = <Byte> 1-9 (XTC/3S ) or 1-32 (XTC/3M)
Parameter Value = <Integer> | <Float>
Integer = 4 byte, low to high
Float = 4 byte, ANSII standard, single precision, low to high
Description: Type of parameter value depends on command ID,
see “DataType / Format” column of Table 5-2.
Response:
None (Just header and trailer)
Special Case: UF255 <Film Number> <List of all Film Parameter Values in
order> will update all film parameters for the given film, in numeric order.
NOTE: Some limitations apply as to when specific film parameters may be
updated, e.g., Hardware used by a film cannot be redefined while that film
is active. All parameters may be updated in Ready state.