Note that the PEEP value will appear on the display:
Caution: PEEP value is influenced by the oxygen enrichment
method. For example, if we connect an Oxygen Mixer for O
2
enrichment, the PEEP may decrease at 3-4 cm H
2
O. Although the
ventilator will issue an alarm, the user may have to readjust the
PEEP by rotating the PEEP knob.