EasyManuals Logo

Inovytec Ventway Sparrow User Manual

Inovytec Ventway Sparrow
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
Ventway Sparrow User Manual
Page 86 of 144
14. CLEANING AND DISINFECTING
Caution: The system is approved for IP45 in operation mode with
oxygen enrichment. To avoid damage to the instrument or
patient
cable, be careful of liquid spillage while cleaning.
Caution: Do not expose the instrument, patient cable or sensors to
sprays, or any other type of solvents.
Caution: Be sure to turn the power off and disconnect the AC power
cord from the power source before performing
cleaning
procedures.
This instrument requires routine cleaning, which includes removal of any
soil or dirt from the external surfaces. A soft cloth dampened lightly with
water may be used.
Spray the entire surface of the device with 70% alcohol.
Leave the alcohol on the device for an exposure time of 2-3 minutes.
Wipe the surface of the device with Pharma-Wipes 70% alcohol.
Inspect the device for residual debris. Should any debris remain, repeat the
entire cleaning procedure until all debris is removed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inovytec Ventway Sparrow and is the answer not in the manual?

Inovytec Ventway Sparrow Specifications

General IconGeneral
BrandInovytec
ModelVentway Sparrow
CategoryMedical Equipment
LanguageEnglish