12.1.2. RECOMMENDED DEVICES FOR MONITORING OF OXYGEN
Company Name
Product Model
Precision Medical
PM5900 Oxygen Monitor
Maxtec
MaxO
2
ME®
ENVITEC
MySign®O
Caution: Use of the low-pressure oxygen system at concen-
trations above 60% is NOT recommended, as higher values
combined with varying minute volume due to spontaneous
breathing of the patient may cause inadvertent PEEP.
Caution: Please read the manufacturer's instructions before using
the oxygen monitoring device.
Caution: Oxygen enrichment delay time until reaching 95%
oxygen at the patient port: 45 [S].
Caution: Measurement of oxygen should be performed using an
external oxygen sensor. The location of the sensor should be at
the outlet of the ventilator between the ventilator and the patient
circuit.