Ventway Sparrow User Manual
Page 44 of 144
If the self test fails, the user is advised to check the following:
• No Oxygen Mixer connected to the ventilator
• No oxygen reservoir
• Dirty/obstructed inlet filter
• Pressure and control tube connected properly (kinks, disconnec-
tion or rupture of the tubes)
• No HME or filter are connected to the exhalation valve
• Patient is not connected
9.3. PATIENT WEIGHT
Rotate the scroll knob to select the weight of the patient:
The patient circuit type will be identified automatically, and the appropriate
weight screen (Pediatric/Adult) will be displayed.
Pediatric and adult weight selection screens
Note: Setting patient weight will automatically set all ventilation