Ventway Sparrow User Manual
Page 44 of 144
If the self test fails, the user is advised to check the following:
No Oxygen Mixer connected to the ventilator
No oxygen reservoir
Dirty/obstructed inlet filter
Pressure and control tube connected properly (kinks, disconnec-
tion or rupture of the tubes)
No HME or filter are connected to the exhalation valve
Patient is not connected
9.3. PATIENT WEIGHT
Rotate the scroll knob to select the weight of the patient:
The patient circuit type will be identified automatically, and the appropriate
weight screen (Pediatric/Adult) will be displayed.
Pediatric and adult weight selection screens
Note: Setting patient weight will automatically set all ventilation
parameters.