23.17. TEMPERATURE ALARM
Step Procedure Action Expected Results
1
Start up the Sparrow
ventilator.
Complete the patient
circuit test.
The software switches to the ventilation phase, weight
selection screen.
2
Set the simulator to
the patient weight of
60 kg.
Select 60 kg and then
SIMV ventilation
mode, Invasive/Non-
Invasive, and confirm
the patient
connection.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Simulate a device
temperature above
MaxAllowedTemper-
ature.
The software indicates that the temperature is above range
(above 80 degrees).
4
Silence the alarm. Alarm indication on the screen, but alarm is silenced.