11.3. SUMMARY OF ALARM LEVELS
Alarms are defined with three levels of importance:
High: Five beeps every two seconds
Medium: Three beeps every five seconds
Low: One beep one time only. Not shown in list of active alarms.
Alarm
category
Alarm
High level
Blower malfunction
Tube disconnect
Patient disconnect
Apnea
System recovered from a
crash
Battery empty
Sensor disconnect
Low respiratory rate
High minute volume
Low minute volume
High inspiratory pressure
Low inspiratory pressure
Leak
Inverse I:E ratio
High temperature
Expiratory valve blocked
High PEEP
Low PEEP
Alarm
category
Alarm
Medium
level
High respiratory rate
Tidal Volume Limit
Reached
Low tidal volume
Low pressure
Battery low
High Voltage
Low level
Replace filter
Service needed
Altitude out of range