7.4. STARTUP DURING VENTILATION
If power is lost during ventilation and the ventilator is turned on within 2
minutes, automatically start with ventilation, and a DISCONNECT alarm of
unexpected shutoff will be displayed on the screen.
Unexpected shutoff screen
If power is lost during ventilation and the ventilator is turned on after
more than 2 minutes, the operator will be asked if he wants to continue
ventilating using the previous parameters.
Unexpected shutoff screen