4.2.2. IMMUNITY
Electromagnetic Environment – Guidance
discharge (ESD)
± 8kV contact
±2, 4, 8, 15 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30 %.
transient / burst
± 2 kV for power
supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC
61000-4-5
mode
± 2 kV common
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
0% .5 Periods
0% 1 Period
70% 25 Periods
0% 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the Ventway Sparrow requires continued
operation during power mains interruptions
beyond that provided by the battery, it is
recommended that the Ventway Sparrow is
powered from an uninterruptible power supply.
Note percentages are the level of mains power
frequency
(50/60 Hz)
magnetic field
30A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
IEC 61000-4-6
Radiated RF IEC
61000-4-3
6 V rms in ISM and
amateur radio bands
10 V/m
And immunity to
proximity fields from
RF wireless
communications
equipment
The Ventway Sparrow is suitable for the
electromagnetic environment of typical
commercial or hospital settings.