EasyManuals Logo

Inovytec Ventway Sparrow User Manual

Inovytec Ventway Sparrow
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #133 background imageLoading...
Page #133 background image
Ventway Sparrow User Manual
Page 133 of 144
23.14. LOW BATTERY ALARM
Step Procedure Action Expected Results
1
Start up the Sparrow
ventilator.
Complete the patient
circuit test.
The software switches to the ventilation phase, weight
selection screen.
2
Set the simulator to
the patient weight of
60 kg. Select 60 kg
and then SIMV
ventilation mode,
Invasive/Non-
Invasive, and confirm
the patient
connection.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Wait until the battery
level reaches
CriticallyLowBattery
level.
The software indicates CriticallyLowBattery level.
4
Silence the alarm. Alarm indication on the screen, but alarm is silenced.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inovytec Ventway Sparrow and is the answer not in the manual?

Inovytec Ventway Sparrow Specifications

General IconGeneral
BrandInovytec
ModelVentway Sparrow
CategoryMedical Equipment
LanguageEnglish