23.14. LOW BATTERY ALARM
Step Procedure Action Expected Results
1
Start up the Sparrow
ventilator.
Complete the patient
circuit test.
The software switches to the ventilation phase, weight
selection screen.
2
Set the simulator to
the patient weight of
60 kg. Select 60 kg
and then SIMV
ventilation mode,
Invasive/Non-
Invasive, and confirm
the patient
connection.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Wait until the battery
level reaches
CriticallyLowBattery
level.
The software indicates CriticallyLowBattery level.
4
Silence the alarm. Alarm indication on the screen, but alarm is silenced.