23.21. VENTILATION DURING STANDBY (DUE TO PATIENT
INSPIRATORY EFFORT)
Step Procedure Action Expected Results
1
Start up the Sparrow
ventilator.
Complete the patient
circuit test.
The software switches to the ventilation phase, weight
selection screen.
2
Set the simulator to
the patient weight of
60 kg.
Select 60 kg and then
SIMV ventilation
mode, Invasive/Non-
Invasive, and confirm
the patient
connection.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Put the device in
standby by stopping
the ventilation in the
menu screen.
The blower stops.
4
Generate 3
inhalation triggers.
The blower starts.
The software switches to the running ventilation screen,
continuing the previous ventilation with the same
parameters.