If the device packaging is not intact, do not use the device.
If the device does not turn on, or is not working correctly, discontinue
use. Refer servicing or replacement to qualified service personnel.
Do not disassemble any part of the system components. This system
is not user-serviceable.
Do not use the equipment if it is not working properly or if it has
suffered any damage, for example, by dropping the equipment or
If the LCD screen is cracked or damaged, check whether the screen
can be used, and if not, do not use the device.
If the power button is damaged or stuck, disconnect the patient from
the device and remove the battery.
If the rotator switch does not allow changing parameters, the device
cannot be used.
The Patient Circuit is single use only. If it is not removed from a new
container, it may have already been used and should not be used.
The Patient Circuit should be changed after 24 hours of continuous
use.
Use only the original Inovytec Patient Circuit.
Confirm that the expiration date, found on the Patient Circuit
packaging bag, has not been reached.
The device should be used under medical supervision.
When using external oxygen enrichment, please note the following:
• When using Oxygen Mixer, the oxygen enrichment level will
reach a minimum of 95%.
• When using reservoir bag, the oxygen enrichment level may
vary depending on oxygen flow rate.
It is the user’s responsibility to retain information about the patient
(by USB connection). Storage capacity may be sufficient for at least