During NIV (Non-Invasive Ventilation) the user should use a
capnograph in order to monitor the CO
2
level of the patient.
Covering the ventilator is prohibited.
Ensure that no Latex or natural rubber parts are in patient pathways.
When adding medication to the gas flowing into the patient by using
an MDI or nebulizer, please position between mask/ETT and exha-
lation valve. Choose NIV set up in order to compensate for additional
volume added by the use of nebulizer.
Do not obstruct the gas intake ports.
Discarded used or unused patient circuit is classified as clinical waste.
As such, the user is responsible for complying with all local and
national regulations regarding discarding of clinical waste.
In order to keep the device waterproof, replace the USB cover in its
After replacing the filter, close the cover screws firmly by hand or
using a tool. Be sure to close completely – if not closed properly,
oxygen enrichment may be affected.
VWSP-100MR and VWSP-900 MR can be used at the specified test
location without negative impact on the image quality of the 3T MR
system.
However, special MR Conditional installation requirements
account in order to keep the device at a safe
distance and acceptable spatial magnetic field gradient isocontour.
Otherwise serious injuries might occur due to the device beco
ming
accidentally a projectile hazard moving towards the magnet bore
entr
ance where high spatial magnetic field gradients exist, causing
magnetically induced displacement forces.
Dispose of this device and used sensors in accordance with local
regulations.