EasyManua.ls Logo

I+ME ACTIA BMS Master 4 - Connection of External Balance Resistors

Default Icon
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Hardware Manual
BMS Master 4 / 4.5
Ref.: IR14417 A 15.01.2018 Page 66/76
Modul
Balance current
cell type
Slave 6 A 2X
20,5 Ohm
U [mV]: 1500 … 2800
I [mA]: 73 … 137
Cell type: LTO cell
Slave 6 A 3X
27,5 Ohm
U [mV]: 2500 … 3600
I [mA]: 91 … 131
Cell type: LFP cell
Slave 6 A 4X
37,5 Ohm
U [mV]: 2800 … 4200
I [mA]: 75 … 112
Cell type: NCA cell
The resistor values are adapted on the voltage range of the cell voltages. Thus an
optimum between maximum balance current and maximum board temperature is reached.
View of a Balance module to connect on the ANA module:
3.6.9.1 CONNECTION OF EXTERNAL BALANCE RESISTORS
All resistor respective cell connections are connected to a 24 pole MicroFit connector. At
this connector it is possible to connect external balance resistors to use higher balance
current max. 600mA. External resistors can be connected of 2 ways.
One connection direct at the cell, one connection with a separate wire at the 24
pole connector.
or
With 2 wires at the 24 pole connector
The cell voltage connections of the 12 pole connector must not use for the external
balance resistors. Wire gauge should be minimum 0.5 mm².
Resistor value is set depending to cell type, so that the balance current lower than

Table of Contents