4230 Flow Meter
Section 1 Introduction
1-8
Optional Stainless steel extension tubes for the bubble line.
4 ft. 1.2 m long × 0.065" (0.17 cm) or 0.125" (0.32 cm) ID
Compensated Temperature
Range
0° to 60°C
Miscellaneous
Flow Rate Calculations The flow meter creates a table of level-versus-flow rate from program selections;
this table divides the level span into 256 equally-spaced level increments. Each
level increment corresponds to a specific flow rate. During data collection, if the
measured level falls between two table values, the flow meter will perform a linear
interpolation to calculate the flow rate value.
Table 1-3 Technical Specifications for FLOWLINK and Bubbler (Continued)
Table 1-4 4230 Chart Longevity
Chart Speed, Inch/Hour
(Note: Report Generator is OFF)
Time to Empty Roll
4 195 Hours (8
1
/8Days)
216
3
/4Days
132
1
/2 Days
0.5 65 Days
Table 1-5 Battery Life Expectancy
Flow Meter Settings Minimum
Default
2
Maximum
Bubble Rate 1.5 / Second 1.5 / Second 1.5 / Second
Reading Interval 5 Minutes Continuous Continuous
Printer Off Off 4" per Hour
Report Generator Off Off Every Hour
Purge Interval 1 Hour 15 Minutes 5 Minutes
Purge Duration
1
/2 Second
1
/2 Second 3 Seconds
Average Current 14 mA 15 mA 27 mA
Nickel-Cadmium
3
10.7 Days
5
10 Days 5.5 Days
Lead-Acid
4