4230 Flow Meter
Section 2 Programming
2-51
Or:
Select YES to reset the sampler enable feature; select NO to
leave the sampler enabled:.
This selection allows you to turn the flow meter's printer on or off
when the sampler is enabled from the flow meter. This allows
you to conserve battery power and print a chart only when the
sampler is enabled. This feature is useful for monitoring storm
water runoff.
2.13 Step 7 - Alarm Dialout
Mode
This step lets you signal an alarm to a remote location from the
flow meter. The conditions that can cause an alarm are the same
as described previously for sampler enabling.
Note
You must have the optional internal modem installed and con-
nected to a telephone network to make use of this feature. This
menu will not even appear if you do not have a modem. The
flow meter will automatically advance to the next step.
More information about the modem can be found in Section 4. If
you have the modem installed and select step 7, the following
will appear:
If you select DISABLE, this option will be inactivated until you
change the selection later. The program will advance to the next
step. If you select STORM, the flow meter will request definitions
similar to those for STORM in sampler enable. The following will
appear:
Followed by:
Then:
ENABLE CURRENTLY LATCHED, RESET
• NO • • YES •
PRINTER ON/OFF WITH ENABLE
• YES • • NO •
ALARM DIAL OUT
• DISABLE • • CONDITIONAL • • STORM • • FLOWLINK •
LEVEL
GREATER THAN X.XXX FT (Or other units of measure.)
RAINFALL AMOUNT
X.XX INCHES (or other units of measure)
RAINFALL TIME PERIOD
• 15 MIN • • 30 MIN • • 1 HR • • 2 HR • • 4 HR •