3-10
KaVo OP 3D Vision Service Manual
032-0358-EN Rev B
9. Select Continue. A message is displayed that the UFD drive will be erased, and the software will
be copied.
10. Select Continue. The UFD is erased, then the copy operation begins.
NOTE: The copy operation takes approximately 30 minutes, depending on the size of the update.
11. When the copy is complete, a success message is displayed. Select Continue.
12. Remove both UFDs from the scanner controller.
13. Select Quit. Installation Progress Monitor is displayed briefly, followed by a message stating no
update was performed.
14. Select . The scanner controller will reboot to the Login screen.
SmartScan STUDIO Manager Settings Window
In addition to providing options for configuring SmartScan STUDIO Manager (described in Installation
Manual), the Settings window also contains the product version and a rescan option.
In SmartScan STUDIO Manager, select to access the Settings window.
Product Version
Starting with version 2.1, the current installed version number of SmartScan STUDIO Manager can be
found on the Settings window.
Rescan
The Rescan option rebuilds the patient exam list database. Use this option if it is suspected that the
exam list is corrupted. Depending on the size of the database, this can be a time-consuming operation.