5-5
IEC Command Codes
032-0358-EN Rev B
X-ray On: Typically equals 1. 0 to (Ticks per frame - 1)
X-ray Off: Greater than X-ray On and less than (Ticks per frame - 1)
Set as needed for desired length. 50,000 = 1 (ms)
Panel Readout: 0 to (Ticks per frame - 1). Typically set the same as
X-ray Off, except for continuous X-ray mode.
Pulse Mode: 0 or 1. If not entered, defaults to 0.
0 = 200 microsecond pulse for panel blanking and valid frame.
1 = either a 5msec pulse for valid frame or 3msec for panel blanking
Tube Range: If not entered, defaults to zero.
0 = 120 KV tube, 1 = 90 KV tube
X-Ray Commands
XCN Performs a cine X-ray acquisition for the number of frames requested at the
‘Ticks Per Frame’ rate as setup in the X-ray register.
Command syntax: XCN Frames Watchdog
Frames: 1 to 1024
Watchdog: 0 = disabled, 1 = enabled
XCT Perform a CT scan for requested distance, speed, and watchdog settings.
Command syntax: XCT Distance Speed Watchdog
Distance: 1000 to 668563
Speed: 1000 to 65000
Watchdog: 0 = disabled (default), 1 = enabled
XIP Performs a Panoramic scan for the requested parameters.
Command syntax: XIP Distance Speed PlatformOffset PlatformStartSpeed
PlatformStopSpeed RampReduction Watchdog
Distance: 1000 to 668563
Speed: 1000 to 70000
Platform Offset: 1 to (Distance - 25000)
Platform Start: 1000 to 54000
Platform End: 1000 to 54000
Ramp Reduction: 0 to 250
Watchdog: 0 = disabled, 1 = enabled
XP Performs an exposed X-ray acquisition for the number of frames requested at the
‘Ticks Per Frame’ rate as setup in the X-ray register.
Command syntax: XP Frames Watchdog
Frames: 1 to 1024
Watchdog: 0 = disabled, 1 = enabled
Miscellaneous Commands
BD Get Board Information
Command syntax: BD
Retrieves the board component versioning.
Results returned in format: OK KRL:x UBT:x FID:x SWB:x
KRL: Linux kernel version (short)
UBT: U-Boot version (short)
FID: 1210 Board Flash (EEPROM) Chip System ID
SWB: ROMFS Software Build version (HG-ID Date Time)

Table of Contents