6-1
032-0358-EN Rev B
Chapter
6
Safety Checks
Recommended Safety Checks
The safety checks in this chapter must be performed on systems installed in locations that must comply with
VDE 0751-1 and/or IEC 62353 standards. They are recommended for all other installations.
Safety checks are required by standard after:
System Repair
Recurrent Maintenance and Interval (2 years)
System Modifications
Initial Installation
Qualified personnel must perform these tests. Qualifications include training on the subject, knowledge,
experience, and acquaintance with the relevant technologies, standards and local regulations. The
personnel assessing the safety must be able to recognize possible consequences and risks arising from
non-conforming equipment.
Responsibility is accepted for the safety, reliability, and performance of the component suppliers by Imaging
Sciences International provided:
Installations, upgrades, adjustments, changes, or repairs are performed by service technicians
trained by ISI or authorized third parties including personnel of authorized distributors.
The system is operated in accordance with the manufacturer’s instructions, including care and
installation instructions.
Safety Tests
The following tests must be performed.
Visual Inspection
Protective Earth Resistance
Insulation Resistance
Earth Leakage Current
Functional Tests