EasyManua.ls Logo

Keithley 2182

Keithley 2182
339 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 ANALOG OUTPUT
Provides a scaled non-inverting DC voltage. With analog output gain set to one, a full range
input will result in a 1V analog output.
2 TRIGGER LINK
Eight-pin micro-DIN connector for sending and receiving trigger pulses among connected
instruments. Use a trigger link cable or adapter, such as Models 8501-1, 8501-2, 8502, and 8503.
3 RS-232
Connector for RS-232 operation. Use a straight-through (not null modem) DB-9 shielded
cable.
4 IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as the Models 7007-1
and 7007-2.
5 Power Module
Contains the AC line receptacle, power line fuse, and line voltage setting. The instrument can
be configured for line voltages of 100V/120V/220V/240VAC at line frequencies of 45Hz to
66Hz or 360Hz to 440Hz.
1-12 Getting Started

Table of Contents

Other manuals for Keithley 2182

Related product manuals