EasyManuals Logo

Keithley 2303 User Manual

Keithley 2303
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Status Structure 7-13
Measurement event status
The used bits of the measurement event register (shown in Figure 7-6) are described as
follows:
Bit B3, reading overflow (ROF) — Set bit indicates that voltage or current reading
exceeds the measurement range of the instrument.
Bit B4, pulse trigger timeout (PTT) — Set bit indicates that a current pulse has
not been detected.
Bit B5, reading available (RAV) — Set bit indicates that a reading was taken and
processed.
Bit B9, buffer full (BF) — Set bit indicates that the specified number of readings
(average count) have been taken.
___
BF
___
RAV
PTT
ROF
___
(B15-B10)
(B9)
(B8-B6)
(B5)
(B4)
(B3)
(B2-B0)
___
BF
___
RAV
PTT
ROF
___
(B15-B10)
(B9)
(B8-B6)
(B5)
(B4)
(B3)
(B2-B0)
___
512
___
16
8
___
(2
9
)
(2
5
)
(2
4
)
(2
3
)
___
BF
___
RAV
PTT
ROF
___
(B15-B10)
(B9)
(B8-B6)
(B5)
(B4)
(B3)
(B2-B0)
OR
&
&
&
&
:CONDition?
[:EVENt]?
To MSB bit
of Status Byte
Register
:ENABle <NRf>
:ENABle?
Decimal Weights
Measurement Event
Enable Register
Measurement
Condition Register
Measurement
Event Register
PTT = Pulse Trigger Timeout
ROF = Reading Overflow
BF = Buffer Full
RAV = Reading Available
& = Logical AND
OR = Logical OR
32
Figure 7-6
Measurement event status

Table of Contents

Other manuals for Keithley 2303

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2303 and is the answer not in the manual?

Keithley 2303 Specifications

General IconGeneral
BrandKeithley
Model2303
CategoryPower Supply
LanguageEnglish

Related product manuals