EasyManua.ls Logo

Keithley 2303 - SCPI Commands - Pulse Current Measurements; SCPI Programming

Keithley 2303
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pulse Current Measurements 3-7
SCPI programming
Table 3-1
SCPI commands — pulse current measurements
Commands Description Default Ref
SENSe
:FUNCtion “PCURrent”
:PCURrent
:AVERage <NRf>
:MODE <name>
:TIME
:AUTO
:HIGH <NRf>
:LOW <NRf>
:AVERage <NRf>
:SYNChronize
[:STATe]
:TLEVel <NRf>
:TLEVel
[:AMPs] <NRf>
:MILLiamp <NRf>
:DELay <NRf>
SENSe subystem:
Select pulse current measurement function.
Pulse current configuration:
Specify average count; 1 to 100, or 1 to 5000 (pulse
current digitization).
Select measurement mode; HIGH, LOW or
AVERage.
Set integration times:
Integration times set automatically.
Specify integration time (in sec) for high pulse
measurements; 33.33e-6 to 0.8333.
Specify integration time (in sec) for low pulse
measurements; 33.33e-6 to 0.8333.
Specify integration time (in sec) for average
pulse measurements; 33.33e-6 to 0.8333.
Pulse detection triggering:
Send ON to select pulse current measurements.
OFF selects pulse current digitization.
Models 2303 and 2303B - Set trigger level in
amps; 0 to 5.
Model 2303-PJ trigger level:
Set trigger level (in amps) for 5A range; 0 to 5.
Set trigger level (in amps) for 500mA range; 0
to 0.5.
Specify trigger delay in seconds; 0 to 0.1 or 0 to 5
(pulse current digitization).
VOLT
1
HIGH
3.333e-5
3.333e-5
3.333e-5
ON
0.0
0.0
0.0
0.0
A
B
C
C
C
D
E
E
F
READ?
READ:ARRay?
Trigger and return one reading.
Trigger an array of readings and return them.
B
B

Table of Contents

Other manuals for Keithley 2303