DISPlay, FORMat, and SYSTem 10-7
:SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in Table
10-3.
A. SYSTem:POSetup <name>Program power-on defaults
Parameters <name> = RST Power-up to *RST defaults
SAV0 Power-up to setup stored in memory location 0
SAV1 Power-up to setup stored in memory location 1
SAV2 Power-up to setup stored in memory location 2
SAV3 Power-up to setup stored in memory location 3
SAV4 Power-up to setup stored in memory location 4
SAV5 Power up to SAV0 (output on or off)
SAV6 Power-up to SAV1 (output on or off)
SAV7 Power-up to SAV2 (output on or off)
SAV8 Power-up to SAV3 (output on or off)
SAV9 Power-up to SAV4 (output on or off)
NOTE SAV5-SAV9 are not available for the Model 2303-PJ.
Table 10-3
SCPI commands — system
Command Description Default Ref
SYSTem
:LFRequency?
:POSetup
<name>
:VERSion?
:ERRor
:CLEar
Read power line frequency.
Select power-on setup: RST or SAVx where:
x = 0 to 4 Model 2303-PJ
x = 0 to 9 Model 2303/2303B
Query SCPI revision level.
Read and clear oldest message in error queue:
Clear messages from error queue.
(see
Note)
Sec 1*
A
Sec 7
Sec 7
* See “Line power connection” under “Power-up” for details.
Note: Clearing the error queue — Power-up and *CLS clears the error queue. *RST and STATus:PRESet have
no effect on the error queue.