10-8 DISPlay, FORMat, and SYSTem
With RST selected, the power supply powers up to the *RST default conditions. Default con-
ditions are listed in the SCPI tables.
With SAV0-4 specied, the power supply powers-on to the setup that is saved in the specied
memory location using the *SAV command (Section 8). Note that the instrument will power up
with the output off.
For the Models 2303 and 2303B, the SAV5-9 parameters can be used to power-up the power
supply with the output on. It will assume the corresponding SAV0-4 setup, including the saved
state of the output. For example, assume the output was on when the setup was saved by *SAV 2.
If you select SAV7 as the power-on setup, the instrument will power-up to the SAV2 setup and the
output will be on. However, if the output was off when setup 2 was saved and power-on setup is
SAV7, then the instrument will power-up with setup 2 values with the output off.