9-4 Signal Oriented Measurement Commands
:MEASure[:<function>]? Execute :READ? on specied function
:MEASure:ARRay[:<function>]? Execute :READ:ARRay? on specied function
Parameters <function> = CURRent[:DC] Measure current
VOLTage[:DC] Measure voltage
PCURrent Measure pulse-current
DVMeter Measure DVM input
LINTegration Perform long integration current
measurements.
When the MEASure? command is sent, the specied function is selected and then the READ?
is executed. When the MEASure:ARRay? command is sent, the specied function is selected
and the READ:ARRay? command is executed. See READ? and READ:ARRay? for details.
If a function is not specied, the measurement(s) will be performed on the function presently
selected.
NOTE There are no AVERage commands for long integration measurements. The array size
for long integration readings is fixed at one. Therefore, MEASure:LINTegration? and
MEASure:ARRay:LINTegration? are basically the same.